The Extech SDL150 is a Dissolved Oxygen Meter with an integrated SD card data logger. This device is designed for precise measurement of dissolved oxygen (DO) concentration and temperature in various applications. It records data directly onto an SD card in Excel format for easy transfer and analysis on a personal computer.
Key features include a dual backlit display, automatic temperature compensation, offset adjustment for relative measurements, and user-programmable sampстапки на линг.

Слика 1: Extech SDL150 Dissolved Oxygen Meter and Probe. This image shows the main unit with its digital display and control buttons, connected to the dissolved oxygen probe.
2. Содржина на пакетот и поставување
2.1. Содржина на пакетот
Verify that all items are present and in good condition upon unpacking. The standard package includes:
- Extech SDL150 Dissolved Oxygen Meter
- Dissolved Oxygen Probe with built-in temperature sensor
- 6 x AA батерии
- SD card (2G capacity)
- Two spare replacement membranes for the DO probe
- Electrolyte solution for the DO probe
- Цврста торбичка за носење
2.2. Почетно поставување
- Инсталација на батерија: Open the battery compartment on the back of the meter and insert the 6 AA batteries, observing correct polarity.
- Вметнување SD-картичка: Locate the SD card slot, typically on the side of the meter. Insert the provided 2G SD card until it clicks into place.
- Probe Preparation:
- Unscrew the protective cap from the dissolved oxygen probe.
- Fill the probe's membrane cap with the supplied electrolyte solution. Ensure no air bubbles are trapped.
- Carefully screw the membrane cap back onto the probe, ensuring the membrane is intact and free of damage.
- Поврзување со сонда: Connect the prepared DO probe to the designated input port on the top of the SDL150 meter.
- Вклучување: Притиснете го МОЌ копче за вклучување на уредот.
3. Упатства за работа
3.1. Основно мерење
- Калибрација: Before first use and periodically thereafter, calibrate the dissolved oxygen probe according to the instructions in the full user manual (not provided here). This typically involves air calibration or zero-oxygen solution calibration.
- Потопување: Immerse the DO probe into the sample solution. Ensure the membrane is fully submerged and free from air bubbles.
- Приказ за читање: The dual backlit display will show the dissolved oxygen concentration (in mg/L or %) and the temperature (in °F or °C).
- Units Selection: Користете го ФУНКЦИЈА button to cycle through different measurement units (mg/L, %, °F, °C) if applicable.

Слика 2: Extech SDL150 in a laboratory environment. The meter is shown measuring dissolved oxygen in a beaker, demonstrating its application.
3.2. Пријавување податоци
- Рачно евидентирање: Притиснете го ДНЕВНИК button to manually store the current reading. The meter can store up to 99 manual readings.
- Автоматско евидентирање:
- Access the settings menu (refer to the full manual for specific button combinations) to set the desired sampстапка на линг.
- Откако сampling rate is set, activate automatic logging. The meter will record up to 20 million readings on the 2G SD card.
- Data is stored in Excel-compatible format for easy transfer to a PC via the SD card.
3.3. Дополнителни функции
- Задржување на податоци: Притиснете го ДРЖИ button to freeze the current display reading. Press again to release.
- Мин/макс: Користете го MIN / MAX копчето за да view the minimum and maximum recorded values during a measurement session.
- Задно осветлување: The display features a backlight for improved visibility in low-light conditions. Activate it using the dedicated backlight button (often indicated by a light bulb icon).
- Автоматско исклучување: The meter includes an auto power-off function to conserve battery life. This feature can typically be disabled or adjusted in the settings.
- Подесување на офсет: The offset adjustment function allows for zeroing the meter for relative measurements. Consult the full manual for detailed instructions on using this feature.
4. Одржување
4.1. Probe Membrane and Electrolyte Replacement
The dissolved oxygen probe requires periodic maintenance, specifically the replacement of the membrane and electrolyte solution. This ensures accurate readings and extends the life of the probe.
- Unscrew the membrane cap from the probe body.
- Carefully remove the old membrane and discard it.
- Rinse the probe tip and the inside of the new membrane cap with distilled water.
- Fill the new membrane cap with fresh electrolyte solution, ensuring no air bubbles are present.
- Gently screw the new membrane cap onto the probe body. Avoid overtightening.
- Allow the probe to stabilize for a few minutes before use and perform a calibration.
4.2. Чистење и складирање
- Чистење на броилата: Избришете ја надворешноста на мерачот со мека, дamp крпа. Не користете абразивни средства за чистење или растворувачи.
- Чистење на сонда: After each use, rinse the probe tip with distilled water. Store the probe with its protective cap on, ideally with a small amount of electrolyte solution to keep the membrane moist, as specified in the full manual.
- Долгорочно складирање: For extended storage, remove the batteries from the meter and ensure the probe is properly prepared as per the full manual's instructions. Store the unit and accessories in the hard carrying case in a cool, dry place.
5 Смена на проблеми
This section provides general guidance for common issues. For detailed troubleshooting steps, refer to the comprehensive user manual provided with your Extech SDL150.
- Без моќност: Проверете ја инсталацијата на батериите и осигурајте се дека батериите не се испразнети. Заменете ги доколку е потребно.
- Неточни читања:
- Изведете калибрација.
- Check the probe membrane for damage or air bubbles. Replace the membrane and electrolyte if needed.
- Ensure the probe is properly connected to the meter.
- SD картичката не снима:
- Verify the SD card is correctly inserted.
- Check if the SD card is full. Transfer data to a PC and clear the card if necessary.
- Ensure the logging function is activated and the sampling rate is set.
6. Спецификации
| Параметар | Вредност |
|---|---|
| Dissolved Oxygen Range | 0 to 20.0 mg/L (0 to 100.0%) |
| Температурен опсег | 0 до 50°C (32 до 122°F) |
| Точност на мерење | +/-0.5% |
| Температурна компензација | Automatic, 0 to 50°C |
| Капацитет за евиденција на податоци | 99 manual readings, 20M readings via 2G SD card |
| Формат на податоци | Excel compatible (.xls) |
| Интерфејс за компјутер | Built-in (via SD card) |
| Извор на енергија | 6 x AA батерии (вклучени) |
| Димензии (Д x Ш x В) | 7.2 x 2.9 x 1.9 инчи (18.3 x 7.4 x 4.8 см) |
| Тежина | 1.61 фунти (0.73 кг) |
7. Гаранција и поддршка
For information regarding warranty coverage, technical support, or service, please contact Extech customer service. Refer to the official Extech website or the documentation included with your product for current contact details and warranty terms.
Производител: Extech